CHLORPROTHIXENE ION-SELECTIVE MEMBRANE ELECTRODES DEVELOPMENT AND APPLICATION IN PHARMACEUTICAL ANALYSIS

Citation
H. Hopkala et al., CHLORPROTHIXENE ION-SELECTIVE MEMBRANE ELECTRODES DEVELOPMENT AND APPLICATION IN PHARMACEUTICAL ANALYSIS, Die Pharmazie, 52(4), 1997, pp. 307-309
Citations number
21
Categorie Soggetti
Pharmacology & Pharmacy",Chemistry
Journal title
ISSN journal
00317144
Volume
52
Issue
4
Year of publication
1997
Pages
307 - 309
Database
ISI
SICI code
0031-7144(1997)52:4<307:CIMEDA>2.0.ZU;2-I
Abstract
The liquid and poly(vinyl chloride) membrane electrodes responsive to the chlorprothixene cation are described. These electrodes are based o n the use of ion-pair complexes of the chlorprothixene picrolonate and chlorprothixene tetraphenylborate. The electrodes exhibit nearly Nems tian response values in different concentration ranges depending on th e used counter ions. The basic analytical parameters of these electrod es were evaluated. The potentiometric titration method was used to det ermine the amount of chlorprothixene hydrochloride in pure form and in tablets with an average of 100,4% and a relative standard deviation o f 1,7%.