L. Gonzalez et al., Report of an image quality and dose audit according to Directive 97/43/Euratom at Spanish private radiodiagnostics facilities, BR J RADIOL, 72(854), 1999, pp. 186-192
Citations number
14
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
An audit of Spanish private medicine radiodiagnostics facilities has been c
arried out, based partly on Spanish legislation relating to European Direct
ives on health protection against ionizing radiation risks in medical expos
ure. The study included an appraisal of infrastructure and equipment, and a
spects of quality assurance and radiation protection, by means of data coll
ected through surveys. Of the 51 centres audited, a sample of 24 X-ray room
s was chosen, then an external evaluation with regard to image quality and
patient dose was performed, by an advisory board of radiologists and medica
l physicists. The methodology used was similar to that of the group of Euro
pean Union experts in European dose evaluation and image quality trials. Ch
est, abdomen, lumbar spine and breast examinations were monitored. Doses we
re measured with thermoluminescent dosimeters. A third of the X-ray rooms e
valuated reached or exceeded dose reference values, and in a third of the c
ases the image quality left considerable room for improvement. Breast and c
hest examinations showed themselves to be the hardest to perform, not only
as a result of exceeding the reference doses, but also due to failure to me
et good image quality standards.