Capacity of uncoated 0.53 mm id pre-columns for retaining sample liquid inthe presence of a solvent vapor exit

Citation
E. Boselli et al., Capacity of uncoated 0.53 mm id pre-columns for retaining sample liquid inthe presence of a solvent vapor exit, HRC-J HIGH, 22(3), 1999, pp. 149-152
Citations number
16
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY
ISSN journal
09356304 → ACNP
Volume
22
Issue
3
Year of publication
1999
Pages
149 - 152
Database
ISI
SICI code
0935-6304(199903)22:3<149:COU0MI>2.0.ZU;2-W
Abstract
0.53 mm i.d. uncoated precolumns of about 10 m in length followed by a solv ent vapor exit have become a standard set-up for large volume on-column inj ection. It went unnoticed, however, that the introduction of a vapor exit r equires two modification of previous working guidelines. First, the capacit y of the precolumn to retain sample liquid is increased by a factor of 2.3- 3 as a result of the around 100 times higher carrier gas flow rate. Secondl y, it must be considered that this gain in retention of liquid is lost agai n upon closure of the exit: as the gas flow rate is reduced to a few mL/min , the layer of the residual sample liquid expands about 2.3-3 times. Hence, closure should occur late, and a section of the precolumn must be assigned for this secondary spreading.