Hc. Trap et Jp. Langenberg, Semi-continuous high speed gas analysis of generated vapors of chemical warfare agents, HRC-J HIGH, 22(3), 1999, pp. 153-158
A method is presented for the continuous analysis of generated vapors of th
e nerve agents soman and sarin and the blistering agent sulfur mustard. By
using a gas sampling valve and a very short (15 cm) column connected to an
on-column injector with a 'standard length' column, the system can either b
e calibrated or used for performing high speed gas analyses. When using a f
lame ionization detector, the detection limit was ca. 100 ppb (ca. 0.5-1.0
mg/m(3)). This technique is applied in inhalation toxicokinetic studies of
nerve agents and mustard gas in the guinea pig.