We have successfully developed and fielded an electromechanically cooled ge
rmanium radiation detector (EMc-HPGe) at Lawrence Livermore National Labora
tory (LLNL). This detector system was designed to provide optimum energy re
solution, long lifetime, and extremely reliable operation for unattended an
d portable applications, For most analytical applications, high purity germ
anium (HPGe) detectors are the standard detectors of choice, providing an u
nsurpassed combination of high energy resolution performance and exceptiona
l detection efficiency. Logistical difficulties associated with providing t
he required liquid nitrogen (LN) for cooling is the primary reason that the
se systems are found mainly in laboratories. The EMC-HPGe detector system d
escribed in this paper successfully provides HPGe detector performance in a
portable instrument that allows for isotopic analysis in the field. It inc
orporates a unique active vibration control system that allows the use of a
Sunpower Stirling cycle cryocooler unit without significant spectral degra
dation from microphonics. All standard isotopic analysis codes, including M
GA and MGA + + [1], GAMANL[2], GRPANL[3] and MGAU[4], typically used with H
PGe detectors can be used with this system with excellent results. Several
national and international Safeguards organisations including the Internati
onal Atomic Energy Agency (IAEA) and U.S, Department of Energy (DOE) have e
xpressed interest in this system. The detector was combined with custom sof
tware and demonstrated as a rapid Field Radiometric Identification System (
FRIS) for the U.S. Customs Service [5]. The European Communities' Safeguard
s Directorate (EURATOM) is field-testing the first Safeguards prototype in
their applications. The EMC-HPGe detector system design, recent application
s, and results will be highlighted. (C) 1999 Elsevier Science B.V. All righ
ts reserved.