Operation of gas proportional scintillation counters in a low charge multiplication regime

Citation
Figm. Borges et al., Operation of gas proportional scintillation counters in a low charge multiplication regime, NUCL INST A, 422(1-3), 1999, pp. 321-325
Citations number
18
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
422
Issue
1-3
Year of publication
1999
Pages
321 - 325
Database
ISI
SICI code
0168-9002(19990211)422:1-3<321:OOGPSC>2.0.ZU;2-U
Abstract
The operation of a xenon filled gas proportional scintillation counter unde r low charge multiplication gains is discussed. It is shown that the best e nergy resolution for X-rays is obtained for a reduced electric field, E/p, of about 8 V cm(-1) Torr(-1), leading to a charge gain (for inter-grid dist ance of 5 mm) of about 1.05, as calculated by a Monte Carlo simulation meth od. A discussion is presented concerning the dependence of the E/p values f or better detector performance, on the inter-grid distance. (C) 1999 Elsevi er Science B.V. All rights reserved.