A fiber-optic attenuator calibration system using high frequency substitution method
Citation
R. Singh et T. Kawakami, A fiber-optic attenuator calibration system using high frequency substitution method, REV SCI INS, 70(3), 1999, pp. 1603-1604
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
SICI code
0034-6748(199903)70:3<1603:AFACSU>2.0.ZU;2-U
Abstract
A fiber-optic attenuator calibration system based on a high frequency subst
itution method has been developed using a 30 MHz precision piston attenuato
r equipped with a laser interferometer as a standard. The uncertainty of at
tenuation calibration for a typical 10 dB attenuator is 0.025 dB at two sta
ndard deviation levels. The resolution of the system is 0.001 dB and the dy
namic range is 25 dB in a single step. (C) 1999 American Institute of Physi
cs. [S0034-6748(99)02503-4].