A. Eustaquio et al., Determination of paracetamol in intact tablets by use of near infrared transmittance spectroscopy, ANALYT CHIM, 383(3), 1999, pp. 283-290
Near infrared transmittance spectroscopy was used to determine the analgesi
c paracetamol in a pharmaceutical preparation commercially available as tab
lets. Spectra were recorded on a dedicated instrument that measures the tra
nsmission of intact tablets over the wavelength range 600-1900 nm. Spectral
data were processed by using two multivariate calibration methods, viz. st
epwise multiple linear regression (SMLR) and partial least-squares regressi
on (PLSR). The analgesic contents provided by the two calibration methods w
ere comparable and differed by less than 1% from the reference (UV spectrop
hotometric) value. The calibration graphs constructed to determine the mean
paracetamol content were used to analyse content uniformity in the tablets
. (C) 1999 Published by Elsevier Science B.V. All rights reserved.