Accurate detection of the spontaneous far-field ventricular signal may
be used to determine the ventricular activation, and hence, the inter
val from atrial stimulus to the ventricular R wave (AR interval) using
a standard atrial pacing lead. This can be useful in developing a phy
siological atrial rate responsive (AAIR) pace-maker and in further imp
roving DDD(R) pacing algorithms. In order to better characterize the a
trial sensed far-field ventricular signal, 200 consecutive patients un
dergoing pacemaker implantation were studied. The amplitude of the far
-field ventricular signal was significantly smaller than that of the a
trial deflection. In all recordings, the slew rate oft he atrial defle
ction was larger than that of the far-field ventricular signal. Subdiv
ision of the recordings by electrode position, pocket location, or QRS
duration on the surface ECG resulted in significantly different signa
l characteristics. The amplitude and slew rate of the far-field ventri
cular signal were significantly smaller in bipolar versus unipolar sen
sing. Atrial sensed far-field ventricular recordings could also be obt
ained in the case of ventricular pacing. Our results indicate that acc
urate sensing of the far-field ventricular signal from an atrial pacin
g lead is conceivable in most patients. The different signal character
istics in relation to parameters, such as electrode position, sensing
mode, and pocket location, may be useful in determining the optimal co
nditions for signal sensing.