LIMITATIONS OF THE ALGORITHM TO RECONFIRM VENTRICULAR-FIBRILLATION INA 3RD-GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR

Authors
Citation
S. Mattke et B. Kramm, LIMITATIONS OF THE ALGORITHM TO RECONFIRM VENTRICULAR-FIBRILLATION INA 3RD-GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR, PACE, 20(4), 1997, pp. 993-994
Citations number
4
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
20
Issue
4
Year of publication
1997
Part
1
Pages
993 - 994
Database
ISI
SICI code
0147-8389(1997)20:4<993:LOTATR>2.0.ZU;2-F
Abstract
We report on inadequate shocks delivered by an ICD for VT converting s pontaneously to AF with slow ventricular response. This represents a c orrect behavior of the reconfirmation algorithm, as the detection of i ntervals longer than the fibrillation detection interval and shorter t han the bradycardia escape interval is required to inhibit the shock.