Application of an amperometric detector for the determination of metals byion-pair high-performance liquid chromatography

Citation
En. Shapovalova et al., Application of an amperometric detector for the determination of metals byion-pair high-performance liquid chromatography, J ANALYT CH, 54(3), 1999, pp. 271-274
Citations number
5
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF ANALYTICAL CHEMISTRY
ISSN journal
10619348 → ACNP
Volume
54
Issue
3
Year of publication
1999
Pages
271 - 274
Database
ISI
SICI code
1061-9348(199903)54:3<271:AOAADF>2.0.ZU;2-2
Abstract
The possibilities for the amperometric detection of Cu(II), Fe(III), and Ni (II) complexes with 4-(2-pyridylazo)resorcinol (PAR) after their separation by chromatography were examined. The optimum conditions for the amperometr ic detection of metals were chosen. It was found that the metals can be qua ntitatively determined with the limits of detection (grams per sample) 4.0 x 10(-9) for Cu(II), 2.9 x 10(-9) for Fe(III), and 1.1 x 10(-9) for Ni(II); RSD = 3-8%. A procedure for determining Cu(II), Fe(III), and Ni(II) in app le juice was developed.