The paper is devoted to an estimation of the true profiles of spectral line
s and equipment influence based on complex experimental line shape profiles
in high-frequency discharge by means of a non-linear multiparameter chi-sq
uare fit. The self-absorption and energy transfer effects are taken into co
nsideration. A good agreement with interferometric data from Hg and Hg-Cd h
igh-frequency electrodeless lamps of both natural and isotopic abundance ar
e achieved. The temperature of radiating atoms and the optical density for
several lines are obtained as a function of the high-frequency oscillator w
orking regime. The confidence regions, standard deviations and correlations
of parameters are obtained using statistical modelling. (C) 1999 Elsevier
Science Ltd. All rights reserved.