A radiochromic film extrapolation method is described for the measurement o
f surface dose from high energy photon beams. Extrapolated central axis ent
rance surface dose using Gafchromic film for a 10 x 10 cm(2) field size is
15% +/- 2% and 13% +/- 2% of D-max for 6 and 10 MV x rays, respectively. Ex
trapolated surface dose for a 30 X 30 cm(2) field with a 10 mm perspex bloc
k tray is 49% +/- 2% and 48% +/- 2% of D-max for 6 and 10 MV beams, respect
ively. All results agree with uncorrected Attix parallel plate ionization c
hamber surface ionization within 4% for the same beam energies and configur
ations. (C) 1999 American Association of Physicists in Medicine. [S0094-240
5(99)01902-1].