Evaluation of the Dade Behring N Latex Cystatin C assay on the Dade Behring Nephelometer II system

Citation
Ej. Erlandsen et al., Evaluation of the Dade Behring N Latex Cystatin C assay on the Dade Behring Nephelometer II system, SC J CL INV, 59(1), 1999, pp. 1-8
Citations number
23
Categorie Soggetti
Research/Laboratory Medicine & Medical Tecnology","Medical Research General Topics
Journal title
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION
ISSN journal
00365513 → ACNP
Volume
59
Issue
1
Year of publication
1999
Pages
1 - 8
Database
ISI
SICI code
0036-5513(199902)59:1<1:EOTDBN>2.0.ZU;2-L
Abstract
The Dade Behring N Latex Cystatin C assay, a particle-enhanced nephelometri c immunoassay for measuring serum cystatin C, was evaluated on the Dade Beh ring Nephelometer II. The assay time was 6 min and the throughput was 75 sa mples per hour. The sample volume was 40 mu L and the measuring range was 0 .25 - 7.90 mg/L. Imprecision studies revealed within-run CVs <1.8% and betw een-run CVs <1.8% in the concentration range 0.87 - 4.63 mg/L. Recovery was 92.4 - 101.3%. Linearity studies showed excellent correlation between the theoretical and obtained values. No interferences were detected from haemog lobin <1.0 mmol/L, bilirubin <512 mu mol/L and Intralipid(R) <20 g/L. Stabi lity of cystatin C in serum was 7 days at temperatures from 20 degrees C to - 20 degrees C and 6 months at - 80 degrees C. Measurements of cystatin C in heparin-plasma and EDTA-plasma did not differ significantly from cystati n C measured in serum. Fifty patient samples run on the Dade Behring Nephel ometer II (y) were compared to the Dako Cystatin C assay (x). The Passing-B ablok regression analysis revealed y=1.105x - 0.340. In conclusion, the Dad e Behring N Latex Cystatin C assay was precise and correlated with the Dako Cystatin C assay.