The results of transcatheter atrial septal defect (ASD) occlusion with 2 di
fferent devices (Sideris adjustable buttoned device vs Amplatzer Septal Occ
luder) were compared in 2 consecutive series of patients. Comparative outco
mes were assessed by whether a device was implanted or not, by complication
s and fluoroscopy time of implantation, and by the incidence of residual sh
unting on transthoracic echocardiography at follow-up. The patient and defe
ct characteristics were similar in both groups. Twenty-eight of 33 Sideris
devices and 37 of 39 Amplatzer devices were implanted, The fluoroscopy time
for the Amplatzer implants was 13.4 minutes (range 8 to 41) compared with
23.7 minutes (range 11 to 60.6) for the Sideris implants (p <0.001). The co
mplete occlusion rate for the Amplatzer device was 93% compared with 44% fo
r the Sideris device at 1 year (p <0.001), In conclusion, the Amplatzer dev
ice produces higher occlusion rates of ASDs with shorter fluoroscopy times,
(C) 1999 by Excerpta Medico, Inc.