Two-way, unfolded three-way and three-mode partial least squares calibration of diode array HPLC chromatograms for the quantitation of low-level pharmaceutical impurities

Citation
Kd. Zissis et al., Two-way, unfolded three-way and three-mode partial least squares calibration of diode array HPLC chromatograms for the quantitation of low-level pharmaceutical impurities, ANALYT CHIM, 384(1), 1999, pp. 71-81
Citations number
23
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
384
Issue
1
Year of publication
1999
Pages
71 - 81
Database
ISI
SICI code
0003-2670(19990322)384:1<71:TUTATP>2.0.ZU;2-9
Abstract
A series of two-way diode array chromatograms were recorded at 0.1-0.5% and 1-5% of 3-hydroxypyridine impurity coeluting with 2-hydroxypyridine, recor ded at pH 4.9 (good resolution) and 5.0 (poor resolution). Four methods for PLS calibration, namely, using summed spectral profiles, summed elution pr ofiles, unfolded three-way PLS and true three-way PLS were applied to the d atasets, both using autopredictions and cross-validation. It was found that it was possible to accurately quantify low levels of impurities. Three-way methods often performed worse than two-way methods using the summed spectr al profile probably due to irreproducibility of elution times. (C) 1999 Pub lished by Elsevier Science B.V. All rights reserved.