Two-way, unfolded three-way and three-mode partial least squares calibration of diode array HPLC chromatograms for the quantitation of low-level pharmaceutical impurities
Kd. Zissis et al., Two-way, unfolded three-way and three-mode partial least squares calibration of diode array HPLC chromatograms for the quantitation of low-level pharmaceutical impurities, ANALYT CHIM, 384(1), 1999, pp. 71-81
A series of two-way diode array chromatograms were recorded at 0.1-0.5% and
1-5% of 3-hydroxypyridine impurity coeluting with 2-hydroxypyridine, recor
ded at pH 4.9 (good resolution) and 5.0 (poor resolution). Four methods for
PLS calibration, namely, using summed spectral profiles, summed elution pr
ofiles, unfolded three-way PLS and true three-way PLS were applied to the d
atasets, both using autopredictions and cross-validation. It was found that
it was possible to accurately quantify low levels of impurities. Three-way
methods often performed worse than two-way methods using the summed spectr
al profile probably due to irreproducibility of elution times. (C) 1999 Pub
lished by Elsevier Science B.V. All rights reserved.