The micro-groove detector

Citation
R. Bellazzini et al., The micro-groove detector, NUCL INST A, 424(2-3), 1999, pp. 444-458
Citations number
15
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
424
Issue
2-3
Year of publication
1999
Pages
444 - 458
Database
ISI
SICI code
0168-9002(19990321)424:2-3<444:TMD>2.0.ZU;2-#
Abstract
We introduce the Micro-Groove Detector (MGD), a new type of two-dimensional position-sensitive gas proportional counter produced using advanced Printe d Circuit Board (PCB) technology. The MGD is based on a thin kapton foil, c lad with gold-plated copper on both sides. An array of micro-strips at a ty pical Ditch of 200 mu m is defined on the top metal layer. Using as a prote ction mask the metal left after the patterning, charge amplifying micro-gro oves are etched into the kapton layer. These end on a second micro-strip pa ttern defined on the bottom metal plane. The two arrays of micro-strips can have an arbitrary relative orientation and so can be used for read-out to obtain 2-D positional information. First results from our systematic assess ment of this device are reported: gas gain > 15 000, rate capability above 10(6) mm(-2) s(-1), energy resolution 22% at 5.4 keV, no significant chargi ng or aging effects up to 5 mC/cm and full primary charge collection effici ency even at high drift fields, (C) 1999 Published by Elsevier Science B.V. All rights reserved.