J. Tschischgale et al., EFFICIENCY CALIBRATION OF A SI(LI) DETECTOR IN THE LOW-ENERGY REGION USING ELECTRON BREMSSTRAHLUNG, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 400(2-3), 1997, pp. 387-391
In the low-energy region the determination of the detector efficiency
is complicated by the fact of nonavailability of certificated calibrat
ion sources. To solve this problem the use of electron bremsstrahlung
for determining the relative photon detection efficiency of a Si(Li) d
etector in the low-energy range 0.5-19 keV is proposed. The calibratio
n is based on energy dispersive measurements of electron bremsstrahlun
g emitted by an X-ray tube. Model spectra are computed from thick-targ
et bremsstrahlung spectra in a semiclassical approximation and from a
physical model for the detector response. Model spectra are fitted to
the measured spectra using a parametric adaptor term. Thereby, the err
or of the determined detector efficiency is approximately 5%.