M. Cheregi et Af. Danet, FLOW-INJECTION DETERMINATION OF CHLORIDE-IONS WITH SPECTROPHOTOMETRICDETECTION, Analytical letters, 30(15), 1997, pp. 2847-2858
A single-line FIA system with a stream-sample splitting device for chl
oride determination is presented. The analytical method is based on th
e Fe3+/Hg(SCN)(2)/Cl- system and the absorbance of the red Fe(SCN)(2+)
species monitored spectrophotometrically at 480 nm. Using a stream-sa
mple splitting device, the recorded signal is composed of two merged p
eaks. Three calibration curves were obtained, once injecting the stand
ard solution series, two using the maximum heights of P-1 and P-2 peak
s and one using the height of T trough. The FIA system showed three li
near responses to the concentration of chloride in the ranges 10-100 p
pm (P-1); 10-500 ppm (P-2) and 20-1000 ppm (T), respectively. Also, it
was capable of detecting chloride ions in different types of water wi
th a throughput of 15 samples h(-1) and the RSD for 240 ppm of Cl- (n=
10) were 1.67% (P-1); 2.38% (P-2) and 1.23% (T), respectively The inte
rference of several ions (commonly found in water) on the FIA outputs
was investigated.