R. Calandrino et al., DETECTION OF SYSTEMATIC-ERRORS IN EXTERNAL RADIOTHERAPY BEFORE TREATMENT DELIVERY, Radiotherapy and oncology, 45(3), 1997, pp. 271-274
The execution of an independent control of monitor units (MU) and dose
distribution calculation, together with a check of the data reported
in the treatment chart is an effective tool in strongly reducing the o
ccurrence of systematic errors before treatment delivery. In this pape
r we report the results of the analysis of 6272 controls (about 5000 p
atients) registered over more than 5 years; 70 serious errors (produci
ng a deviation larger than 5% from the prescribed daily dose) and 147
minor errors were detected and corrected before the start of the treat
ment. The error rate was found to be strongly operator-dependent (seri
ous error rate ranging from 0.3 to 2.5% when considering different ope
rators). A time-trend analysis showed a significant reduction of serio
us errors, i.e. 1.5% in the period from September 1991 to April 1994 c
ompared to 0.9% in the period from April 1994 to November 1996. Howeve
r, even if the double check was highly effective in revealing human er
rors, three serious systematic errors (errors occurring during the cal
culation/planning/transcription phases) escaped the control and were d
etected by diode in vivo dosimetry during the period October 1994 to N
ovember 1996 (in 650 patients controlled). (C) 1997 Elsevier Science I
reland Ltd.