A. Bucker et al., MR-GUIDED BIOPSIES WITH AN ULTRAFAST HIGH -RESOLUTION T-2-WEIGHTED TURBO SPIN-ECHO SEQUENCE LOLO - FIRST CLINICAL-RESULTS, RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren, 167(5), 1997, pp. 491-495
Citations number
15
Journal title
RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren
Purpose: The feasability of the ''LoLo''-technique for MR guidance of
biopsy procedures was tested. Material and Methods: MR-guided biopsies
were performed on 10 patients employing a 1.5 T system, The ''Lolo''-
technique used is a single shot turbo spin echo technique. Only a smal
l field of view is covered in order to yield images with a resolution
of 1 mm(2) in 600 ms. The orthogonal orientation of the slice selectiv
e radio frequency pulses to each other prevents foldover artifacts. Re
sults: No complications occurred. All biopsy procedures yielded suffic
ient material to diagnose the underlying disease. The ''LoLo''-techniq
ue enabled good depiction of the needle tip in all cases. T-2-weighted
contrast typical for turbo spin echo images was observed. No foldover
artifacts were detectable. Conclusion: MR-guided biopsies are possibl
e with the ''LoLo''-technique. Compared to gradient echo sequences T-2
-weighting and smaller susceptibility artifacts proved to be advantage
ous.