Ga. Harff et al., EVALUATION OF THE RADIOMETER WHOLE-BLOOD GLUCOSE MEASURING SYSTEM, EML-105, European journal of clinical chemistry and clinical biochemistry, 35(3), 1997, pp. 241-242
The performance of a new glucose electrode system from Radiometer was
tested using two EML(TM)1O5 analyzers (Radiometer Medical AIS, Copenha
gen, Denmark). Results were very precise (both analyzers reported CV=1
.0% at a glucose concentration of 13.4 mmol/l). Comparison of methods
was performed according to the NCCLS EP9-T guideline. Patients glucose
results from both analyzers were lower compared with the results obta
ined with a Hitachi 911 (Boehringer Mannheim, Mannheim, Germany). Ther
e was no haematocrit dependency of relevance.