V. Weldon et al., OXYGEN SENSING USING SINGLE-FREQUENCY DFB LASER-DIODES ART LAMBDA = 761 NM, Sensors and actuators. B, Chemical, 42(3), 1997, pp. 163-168
Oxygen O-2 sensing using high-resolution direct optical absorption wit
h a single mode GaAs-AlGaAs distributed feedback (DFB) laser diode emi
tting at a wavelength around 761 nm which coincides with the b(1) Sigm
a(g)(+) (v' = 0) <-- X-3 Sigma(g)(-) (v '' = 0) electronic transition
band is reported. In particular, issues relevant to O-2 sensing in the
high concentration regime, not previously of concern for trace gas se
nsing, are investigated. While detectivity of 20 ppm over a 1 m pathle
ngth is determined, the need for careful consideration of parameters s
uch as linearity, dynamic range and calibration and the effect of pres
sure-induced broadening on sensitivity is highlighted. (C) 1997 Elsevi
er Science S.A.