Dn. Weiss et al., PREDISCHARGE ARRHYTHMIA INDUCTION TESTING OF IMPLANTABLE DEFIBRILLATORS MAY BE UNNECESSARY IN SELECTED CASES, The American journal of cardiology, 80(12), 1997, pp. 1562-1565
Complete postoperative evaluation of implantable cardioverter-defibril
lators (ICDs) before discharge, including arrhythmia induction, has be
en the standard since their introduction, Whereas the original ICDs pr
ovided little telemetered information and used separate pace-sense and
defibrillation leads, modern, third-generation devices provide pace-s
ense function information in addition to other data and are used in co
njunction with integrated transvenous endocardial leads that combine p
ace-sense and defibrillation function. Changes in lead position, which
can potentially result in either an inability to detect fibrillation
or to terminate it, should be mirrored by changes in resting pace-sens
e function. Thus, for newer ICDs implanted with integrated endocardial
lead systems, it is possible that in at least some cases predischarge
arrhythmia inductions can be avoided. Two hundred patients receiving
third-generation ICDs in conjunction with integrated transvenous leads
were evaluated before discharge. Defibrillation detection or terminat
ion problems were seen in 8. Declines in resting R-wave amplitude and
pacing impedance were significantly associated with such complications
(-7 +/- 5 vs -0.3 +/- 2.3 mV [p <0.0001] and -158 +/- 138 vs -93 +/-
76 Omega [p <0.05], for those with vs without complications, respectiv
ely), as were gross right ventricular lead migrations on chest x-ray.
No patient with a defibrillation complication had an R-wave change of
<3 mV. However, 13% of patients without complications had R-wave chang
es of >3 mV. It is concluded that a pace-sense evaluation of ICDs may
be a satisfactory screen to determine those who need to go on to compl
ete testing with arrhythmia induction in selected cases. (C) 1997 by E
xcerpta Medico, Inc.