Ys. Tsyganov, PARAMETRIZATION OF SHALLOW IMPLANTED NUCLEI ALPHA-DECAY SPECTRA MEASURED WITH A TOTALLY DEPLETED SILICON DETECTOR, Applied radiation and isotopes, 48(9), 1997, pp. 1211-1213
Citations number
6
Categorie Soggetti
Nuclear Sciences & Tecnology","Radiology,Nuclear Medicine & Medical Imaging","Chemistry Inorganic & Nuclear
A computer simulation of a shallow implanted nuclide alpha-decay spect
rum is reported. A 2-dimensional parametrization procedure is proposed
. Comparison is made with a spectrum of Po-211 nuclei produced in a he
avy-ion-induced nuclear reaction measured with a totally depleted (ove
rdepletion ratio = 1.2) silicon radiation detector. Particles detectio
n was performed from the rear contact of the detector. (C) 1997 Elsevi
er Science Ltd. All rights reserved.