ELISA FOR DETECTING OKADAIC ACID IN MODEL SYSTEMS USING PURIFIED POLYCLONAL ANTIBODIES

Citation
Lg. Dixon et al., ELISA FOR DETECTING OKADAIC ACID IN MODEL SYSTEMS USING PURIFIED POLYCLONAL ANTIBODIES, Journal of rapid methods and automation in microbiology, 5(4), 1997, pp. 269-283
Citations number
38
ISSN journal
10603999
Volume
5
Issue
4
Year of publication
1997
Pages
269 - 283
Database
ISI
SICI code
1060-3999(1997)5:4<269:EFDOAI>2.0.ZU;2-L
Abstract
Polyclonal antibodies were produced in rabbits against okadaic acid (O A) following its coupling to bovine serum albumin using standard carbo diimide condensation procedure to form the immunogen. The immunogen wa s dialyzed against Tris buffer pH 7.45 at 4C and used to immunize two rabbits. Each rabbit received (at four sites) 0.5 mt immunogen (i.d.) together with adjuvant followed by three repeated injections of 0.3 mt emulsified immunogen-adjuvant mixture at weekly intervals. The last i njection was made 60 days after the fourth. The antiserum was collecte d, at intervals, the immunoglobulin fraction (IgG) isolated purified a nd used in an ELISA system to capture the okadaic moiety of the immuno gen. Antibody titers increased following repeated immunization and the IgG recognized low levels of OA. However, ELISA was more sensitive fo r detecting the immunogen than for pure OA, but using the biotinylated -IgG enhanced the titration to both and the detection limit for OA war ; 0.63 ng per 0.1 mL buffer-methanol and the assay linearity ranged fr om 0.63 to 5.0 ng OA.