A combined flow injection (FI)-capillary electrophoresis (CE) system w
ith a split-flow interface reported previously was used for on-line pr
econcentration of pseudoephedrine by solid phase extraction on a micro
-column packed with C-18 Aqueous samples were loaded on the column at
1.0 mi min(-1) for 90 s and eluted by a mixed eluent of 40% 62.5 mM ac
etate buffer (pH 4.0) and 60% acetonitrile at 0.2 mi min(-1). A zone-s
ampling approach was employed to sample 45 mu l of the most concentrat
ed zone of the eluate, which was injected into the FI-CE split-flow in
terface using a phosphate buffer carrier, with a 25 mu l water zone pr
eceding the sampled eluate zone to achieve enhanced field amplificatio
n effects. Sensitivity enhancements of 60 fold (180 fold with 4 min sa
mple loading) were thus achieved with a sample throughput of 9 h(-1) a
nd 4.7% relative standard deviations (n = 10, 2 mu g ml(-1) pseudoephe
drine) for peak area evaluation, without significant degradation of co
lumn efficiency. (C) 1997 Elsevier Science B.V.