USEFULNESS OF A RUGGEDNESS TEST IN THE VALIDATION OF FLOW-INJECTION ANALYSIS SYSTEMS

Citation
Lmbc. Alvaresribeiro et Aasc. Machado, USEFULNESS OF A RUGGEDNESS TEST IN THE VALIDATION OF FLOW-INJECTION ANALYSIS SYSTEMS, Analytica chimica acta, 355(2-3), 1997, pp. 195-201
Citations number
22
Journal title
ISSN journal
00032670
Volume
355
Issue
2-3
Year of publication
1997
Pages
195 - 201
Database
ISI
SICI code
0003-2670(1997)355:2-3<195:UOARTI>2.0.ZU;2-R
Abstract
The usefulness of a ruggedness test to evaluate the influence of small variations in the parameters fixed in the assemblage of a flow inject ion analysis (FIA) system on its response is presented. The purpose of the tested system was the determination of calcium in water with incr eased sensitivity by differential potentiometry involving two identica l calcium ion selective electrodes. Youden and Steiner's ruggedness te st was found to be a useful technique to ascertain that the optimized system worked properly. (C) 1997 Elsevier Science B.V.