LINEARITY DISCONTINUITIES IN XE-FILLED X-RAY MICROSTRIP DETECTORS

Citation
G. Zavattini et al., LINEARITY DISCONTINUITIES IN XE-FILLED X-RAY MICROSTRIP DETECTORS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 401(2-3), 1997, pp. 206-216
Citations number
19
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
401
Issue
2-3
Year of publication
1997
Pages
206 - 216
Database
ISI
SICI code
0168-9002(1997)401:2-3<206:LDIXXM>2.0.ZU;2-L
Abstract
A prototype Xe + 10% CH4 microstrip detector was used to study the K-e dge discontinuity in the pulse-height distribution as a function of th e energy of incident X-rays, The electronics used was such that a puls e-shape rejection could be made of K-fluorescence reabsorption in the detector, The measured jump at the K-edge was (169 +/- 12)eV. It was a lso seen that K-alpha- and K-beta-decay channels produce different pul se heights for the same deposited energy. We find that the difference in energy between the calibration fines of the K-beta 2- and the K-bet a 1.3-escape peaks is (93+/-12)eV whereas between the K-beta 1.3 and t he K-alpha-escape peaks' calibration tile energy gal? is (69+/-7)eV.