PH ELECTRODE PERFORMANCE UNDER AUTOMATED MANAGEMENT CONDITIONS

Citation
Jea. Comer et C. Hibbert, PH ELECTRODE PERFORMANCE UNDER AUTOMATED MANAGEMENT CONDITIONS, Journal of automatic chemistry, 19(6), 1997, pp. 213-224
Citations number
8
Categorie Soggetti
Instument & Instrumentation","Biochemical Research Methods
ISSN journal
01420453
Volume
19
Issue
6
Year of publication
1997
Pages
213 - 224
Database
ISI
SICI code
0142-0453(1997)19:6<213:PEPUAM>2.0.ZU;2-L
Abstract
PH is frequently measured in laboratories, but to have confidence in t he results it is necessary to know that it was measured properly. For an electrode to give accurate results it must be treated well and cali brated correctly. In this Paper, an automated system for PH measuremen t is described; the system uses the operational pH scale and calibrate s using two or three buffer solutions, taking proper account of the ef fects of temperature on the system. The system can be programmed with standard methods and procedures to ensure that the electrode gives the best possible performance. Calibrations and measurements within the s ystem are reproducible, and the automated system is more robust than t he manual pH meter, and requires less operator time.