AUTOMATED ONLINE SOLID-PHASE EXTRACTION GAS-CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION - DETERMINATION OF BENZODIAZEPINES IN HUMAN PLASMA

Citation
Ajh. Louter et al., AUTOMATED ONLINE SOLID-PHASE EXTRACTION GAS-CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION - DETERMINATION OF BENZODIAZEPINES IN HUMAN PLASMA, Journal of chromatography B. Biomedical sciences and applications, 689(1), 1997, pp. 35-43
Citations number
16
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
ISSN journal
13872273
Volume
689
Issue
1
Year of publication
1997
Pages
35 - 43
Database
ISI
SICI code
0378-4347(1997)689:1<35:AOSEGW>2.0.ZU;2-2
Abstract
An automated sample preparation module, the ASPEC (automated sample pr eparation with extraction columns) was interfaced with a capillary gas chromatograph (GC) by means of a loop-type interface. The system was optimized for the determination of four benzodiazepines in plasma. Ext raction from the untreated plasma was carried out on disposable C-18 c artridges and involved several washing steps. The analytes were desorb ed with 2 ml of ethyl acetate and a 110-mu l aliquot of the eluate was injected into the gas chromatograph via the loop-type interface using fully concurrent solvent evaporation conditions. Detection of the ben zodiazepines was carried out with a nitrogen-phosphorus detector (NPD) . The ASPEC-GC-NPD was fully automated and could run unattended overni ght. With a sample volume of 1 ml the procedure showed good linearity and repeatability in the range 5-50 ng/ml using a sample volume of 1 m l. The limits of detection in plasma were 0.5-2 ng/ml.