In routine dosimetry we assume separability of the collimator (S-c) an
d phantom (S-p) scatter components that together comprise the total sc
atter factor(S-c.p) In practice, the addition of blocking also affects
the photon fluence attributable to the treatment head and flattening
filter in a complicated way. The reduced aperture blocks out some of t
he head scatter contribution, while the block and tray add back second
ary scatter. In the following we present techniques for directly measu
ring the aperture effect on S-c in air or in a full-scatter phantom. T
he change in S-c is found to be a scaleable quantity that can be model
led as a simple linear fit to the ratio of projected open-to-blocked e
quivalent square fields. Measurements have been made for 6, 18 and 24
MV photon beams on one Varian 2500 and two Varian 2100c accelerators.
Results indicate a progressive loss of collimator scatter contribution
with increased field blocking that is amplified with increasing energ
y. Block and tray scatter only contribute significantly to S-c for lar
ge fields and treatment distances of 80 cm or less. Application of the
se corrections in monitor unit calculations is presented.