HIGH-PRESSURE XENON ELECTRONICALLY COLLIMATED CAMERA FOR LOW-ENERGY GAMMA-RAY IMAGING

Citation
Ai. Bolozdynya et al., HIGH-PRESSURE XENON ELECTRONICALLY COLLIMATED CAMERA FOR LOW-ENERGY GAMMA-RAY IMAGING, IEEE transactions on nuclear science, 44(6), 1997, pp. 2408-2414
Citations number
24
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
44
Issue
6
Year of publication
1997
Part
2
Pages
2408 - 2414
Database
ISI
SICI code
0018-9499(1997)44:6<2408:HXECCF>2.0.ZU;2-7
Abstract
A high pressure xenon scintillation drift chamber with 30 cm diameter field of view has been built and tested as a two-dimensional imaging c amera for low energy gamma-rays. Experiments demonstrate that the came ra can be also operated in an electronically collimated mode as a thre e-dimensional imaging Compton camera. A spatial resolution of 25 mm is measured for a 140 keV point-like source placed 10.5 cm in front of t he camera without any collimator. The resolution is limited by the int rinsic resolution of the 19 channel readout system which is based on 8 0 mm diameter phototubes. A way to improve the imaging performance is discussed.