Y. Narita et al., MONTE-CARLO EVALUATION OF ACCURACY AND NOISE PROPERTIES OF 2 SCATTER CORRECTION METHODS FOR TL-201 CARDIAC SPECT, IEEE transactions on nuclear science, 44(6), 1997, pp. 2465-2472
Two independent scatter correction techniques, transmission dependent
convolution subtraction (TDCS) and triple-energy window (TEW) method,
were evaluated in terms of quantitative accuracy and noise properties
using Monte Carlo simulation (EGS4). Emission projections (primary, sc
atter and scatter plus primary) were simulated for three numerical pha
ntoms for Tl-201. Data were reconstructed with ordered-subset EM algor
ithm including noise-less transmission data based attenuation correcti
on. Accuracy of TDCS and TEW scatter corrections were assessed by comp
arison with simulated true primary data The uniform cylindrical phanto
m simulation demonstrated better quantitative accuracy with TDCS than
with TEW (-2.0% vs 16.7%) and better S/N (6.48 vs 5.05). A uniform rin
g myocardial phantom simulation demonstrated better homogeneity with T
DCS than TEW in the myocardium; i.e., anterior-to-posterior wall count
ratios were 0.99 and 0.76 with TDCS and TEW, respectively. For the MC
AT phantom, TDCS provided good visual and quantitative agreement with
simulated true primary image without noticeably increasing the noise a
fter scatter correction. Overall TDCS proved to be more accurate and l
ess noisy than TEW, facilitating quantitative assessment of physiologi
cal functions with SPECT.