ELECTRODE MICRODISLODGMENT AS A CAUSE OF PSEUDOCAPTURE IN A VDD PACEMAKER

Citation
Er. Font et al., ELECTRODE MICRODISLODGMENT AS A CAUSE OF PSEUDOCAPTURE IN A VDD PACEMAKER, Revista espanola de cardiologia, 50(12), 1997, pp. 909-912
Citations number
6
ISSN journal
03008932
Volume
50
Issue
12
Year of publication
1997
Pages
909 - 912
Database
ISI
SICI code
0300-8932(1997)50:12<909:EMAACO>2.0.ZU;2-#
Abstract
The new AV sequential pacemakers have improved the suitability for the election of the best pacing mode for each patient, The complexity of the systems may mask some dysfunctions. In the presented case, a failu re to capture due to microdislodgment, may have been missed in a simpl e pacemaker control, because of the combination of several factors: th e presence of normal AV conduction at that moment, the concordance bet ween the pacemaker stimulus and the conduced QRS complex and the simil ar morphology of the conduced and paced QRS complex.