H. Iwase et I. Ono, DETERMINATION OF ASCORBIC-ACID IN SOUPS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH ELECTROCHEMICAL DETECTION, Journal of agricultural and food chemistry, 45(12), 1997, pp. 4664-4667
Determination of ascorbic acid (AA) in soups was performed by high-per
formance liquid chromatography with electrochemical detection set at 4
00 mV vs an Ag/AgCl reference electrode. This method was suitable for
the determination of AA in soups. The proposed method was simple, rapi
d (analysis time similar to 8 min), sensitive [detection limit similar
to 0.1 ng per injection (20 mu L) (at a signal-to-noise ratio of 3:1)
], highly selective, and reproducible [relative standard deviation sim
ilar to 3.1% (n = 5)]. The calibration graph of AA was linear in the r
ange of 0-16 ng per injection (20 mu L). Recovery of AA was >90% by th
e standard addition method. Periodical changes of AA in soup were also
examined. it was found that AA standards diluted with both tap water
and hot tap water were very unstable, and AA in soup prepared with hot
tap water (>90 degrees C) according to the label was very stable.