M. Sadowski et al., CALIBRATION OF CN AND CR-39 TRACK DETECTORS FOR MEASUREMENTS OF FAST DEUTERONS AND NITROGEN-IONS, Radiation measurements, 28(1-6), 1997, pp. 201-206
The paper reports on the analysis of responses of CN-type films and CR
39-plastic nuclear track detectors (NTDs) to fast deuterons and nitrog
en ions emitted from high-current Plasma-Focus (PF) facilities, as ope
rated at SINS and INFIP. In order to separate the deuterons and nitrog
en ions of different energy, the use was made of a Thomson-type mass-s
pectrometer adapted to measurements within dense plasma streams. Deute
ron and ion parabolae were registered with the CN-80 and CN-LR115 film
s as well as with the CR39 plastic NTDs, which were undergone routine
etching procedures. On the basis of known characteristics of the spect
rometer applied energy-and mass-scales for the obtained ion parabolae
were determined. Using an optical microscope, a detailed analysis of t
rack dimensions was performed in chosen points along the registered pa
rabolae, and the calibration diagrams (i..e, a track diameter for the
chosen ion energy and different etching times) were determined. Result
s of the calibration procedure, after comparing with other calibration
data obtained with conventional particle accelerators, may be applica
ble for measurements of fast deuterons (i.g., in nuclear fusion experi
ments) and energetic nitrogen ions (i.g., in some plasma-and ion-facil
ities used for technological applications).