A controlled radon chamber was used to test the detection performance
of a portable active radon detector. The elapsed time to reach detecti
on equilibrium with the radon atmosphere was three hours; the radon de
tector was sensitive to daily temperature changes. After using the rad
on detector three hours waiting time is required to place it in a new
radon environment.