DEVELOPING THE SELF-CONTAINED HYDROGEN REFERENCE ELECTRODE

Authors
Citation
Sl. Gong et al., DEVELOPING THE SELF-CONTAINED HYDROGEN REFERENCE ELECTRODE, Journal of electroanalytical chemistry [1992], 436(1-2), 1997, pp. 291-293
Citations number
3
Journal title
Journal of electroanalytical chemistry [1992]
ISSN journal
15726657 → ACNP
Volume
436
Issue
1-2
Year of publication
1997
Pages
291 - 293
Database
ISI
SICI code
Abstract
A new design of self-contained hydrogen reference electrode is reporte d. The filling of electrolyte and hydrogen for the electrode can be ea sily completed in a few minutes without the need for vacuum operation. The hydrogen filling can either be in situ electrolysis or direct gas filling from an external gas source. The potential error is below 3 m V, in most cases below 1 mV, for relatively concentrated acid or alkal ine solutions. The electrode can work normally for about a week for a single hydrogen filling. Besides, this electrode can be moved around a s conveniently as a calomel electrode. This reference electrode should find wide applications in electrochemical experiments. (C) 1997 Elsev ier Science S.A.