De. Weiss et al., DOSE DISTRIBUTIONS IN TUBING IRRADIATED BY ELECTRON-BEAM - MONTE-CARLO SIMULATION AND MEASUREMENT, Radiation physics and chemistry, 50(5), 1997, pp. 475-485
A set of nine tube sleeves was prepared by wrapping polyethylene film
around a mandrel. Wall thicknesses ranged from a nominal 0.3 to 0.6 cm
and diameters ranged from 2 to 5 cm. The sleeves were irradiated with
a scanning 3 MeV electron beam and were dose-mapped by indexing and u
nwrapping the tube sleeves and measuring the dose by an FTIR spectrosc
opic technique. A Monte Carlo simulation model is also described which
includes all important elements of 3-D geometry necessary to describe
this experiment including window foil and beam scan angle. The simula
tion results were compared with the highly detailed dose distribution
measurements in the tubing and the differences were found to be genera
lly within 3 to 6 percent. Published by Elsevier Science Ltd.