E. Bulska et al., ONLINE PRECONCENTRATION OF LEAD AND CADMIUM FOR FLAME ATOMIC-ABSORPTION SPECTROMETRY USING A FLOW-THROUGH ELECTROCHEMICAL MICROCELL, Analytica chimica acta, 357(1-2), 1997, pp. 133-140
A method was developed for the trace determination of lead and cadmium
by flame atomic absorption spectrometry (FAAS) with electrochemical p
reconcentration. Both metals were deposited on the glassy-carbon elect
rode in a flow-through microcell of volume 3 mu l. At a consumption of
a 3 mu I sample volume, the procedure resulted in an enhancement fact
or of more than 20 and yielded a detection limit of 1 mu g l(-1) for P
b and 0.08 mu g l(-1) for Cd. The relative standard deviation (n=12) w
as 2.1% at the 100 mu g l(-1) level for Pb and 3.2% at the 8 mu g l(-1
) level for Cd. Results obtained for the standard reference material,
water SRM 1643c, were in good agreement with the certified values. (C)
1997 Elsevier Science B.V.