E. Benit et al., CORONARY EMBOLIZATION OF THE MEMBRANE FIXING A MULTILINK STENT TO ITSDELIVERY BALLOON, Catheterization and cardiovascular diagnosis, 43(1), 1998, pp. 68-70
We report on a case of coronary embolization of an elastic membrane fi
xing the ACS RX Multilink stent over its balloon, after a successful s
tent delivery. The membrane was dislodged from the balloon in the osti
um of the right coronary artery as the delivery balloon was being pull
ed back into the guiding catheter. All attempts to retrieve the membra
ne, to cover it with another stent, or to push it into a terminal segm
ent of the artery to limit the jeopardized myocardial mass were unsucc
essful. No CABG was performed because the left coronary system was nor
mal and the procedure occurred 2 months after an incomplete inferior m
yocardial infarction. There was no increase in cardiac enzymes nor ele
ctrocardiographic signs of a new myocardial infarction at discharge or
at 1-mo follow-up. It seems preferable to avoid further use of stents
tied to their balloons with an intermediary, and possibly detachable,
element. (C) 1998 Wiley-Liss, Inc.