U. Harborn et al., EVALUATION OF A MINIATURIZED THERMAL BIOSENSOR FOR THE DETERMINATION OF GLUCOSE IN WHOLE-BLOOD, Clinica chimica acta, 267(2), 1997, pp. 225-237
A miniaturized thermal biosensor has been evaluated as part of a flow-
injection analysis system for the determination of glucose in whole bl
ood. Glucose was determined by measuring the heat evolved when samples
containing glucose passed through a small column with immobilized glu
cose oxidase and catalase. Samples of whole blood (1 mu l) can be meas
ured directly, without any pretreatment. The correlation in the respon
se between the thermal biosensor, the Reflolux S meter (Boehringer Man
nheim), the Granutest 100 glucose test kit (Merck Diagnostica) and the
Ektachem (Kodak) instrument was evaluated. The influence of the hemat
ocrit value and of possible interferences is reported. The correlation
measurements show that the thermal biosensor calibrated with aqueous
glucose standards generally gives lower values on blood glucose than t
he reference methods calibrated for serum or blood measurements. Mean
negative biases range from 0.53 to 1.16 mmol/l. Differences in sample
treatment clearly complicate comparisons and the proper choice of refe
rence method. There was no influence from substances such as ascorbic
acid (0.11 mmol/l), uric acid (0.48 mmol/l), urea (4.3 mmol/l) and ace
taminophen (0.17 mmol/l) on the response to 5 mmol/l glucose. The hema
tocrit value does not influence the glucose determination, for hematoc
rit values of between 13 and 53%. (C) 1997 Elsevier Science B.V.