M. Shimura et al., CORRELATION BETWEEN MEASUREMENTS USING A NEW HALITOSIS MONITOR AND ORGANOLEPTIC ASSESSMENT, Journal of periodontology, 68(12), 1997, pp. 1182-1185
HALITOSIS IS KNOWN AS UNPLEASANT Oral Odor and is a health concern amo
ng the general public. Previously, we reported on a new portable monit
or with a zinc-oxide, thin-film, semiconductor sensor which demonstrat
ed simplicity of handling, high reproducibility and correspondence for
organoleptic assessment. The results suggested its usefulness for the
diagnosis of halitosis. Using the monitor, oral air samples of 94 sub
jects were measured in a field survey, and the values were compared wi
th the organoleptic rates of corresponding samples assessed by two jud
ges. A highly significant correlation (r=0.824, P<0.01) was demonstrat
ed between the measures obtained by the two methods. The results sugge
st that the monitor is useful for not only a clinical study but also a
field study of halitosis.