DURABILITY OF OSCILLATING BONE SAW BLADES

Citation
Tw. Ark et al., DURABILITY OF OSCILLATING BONE SAW BLADES, Journal of long-term effects of medical implants, 7(3-4), 1997, pp. 271-278
Citations number
3
ISSN journal
10506934
Volume
7
Issue
3-4
Year of publication
1997
Pages
271 - 278
Database
ISI
SICI code
1050-6934(1997)7:3-4<271:DOOBSB>2.0.ZU;2-D
Abstract
The purpose of this study was to measure the durability of disposable saw blades using a standard bone-cutting experimental model that can b e replicated in any laboratory. Using a constant cutting force, the bo ne-cutting efficiency was determined by recording the maximum cutting temperature and the rate of material removal. In this experimental stu dy, porcine femurs as well as a synthetic bone substitute were used as the test samples for measuring cutting efficiency. Repeated cutting o f these test materials did not significantly alter the performance of the disposable bone-cutting blades.