CALIBRATION OF A 10 1 RATIO TRANSFORMER USING THOMPSONS METHOD/

Citation
Y. Nakamura et al., CALIBRATION OF A 10 1 RATIO TRANSFORMER USING THOMPSONS METHOD/, Metrologia, 34(4), 1997, pp. 353-355
Citations number
8
Journal title
ISSN journal
00261394
Volume
34
Issue
4
Year of publication
1997
Pages
353 - 355
Database
ISI
SICI code
0026-1394(1997)34:4<353:COA11R>2.0.ZU;2-T
Abstract
The results are reported from the calibration, using Thompson's method , of a 10:1 ratio transformer which will be used for setting up scales of capacitance at the Electrotechnical Laboratory, Tsukuba. The devia tion of the ratio from the nominal value of 10:1 was measured as -1, 3 7 x 10(-7) for the in-phase component and -7, 22 x 10(-7) for the quad rature component, with the transformer operated at a frequency of 1592 Hz with an rms applied voltage of 100 V. The uncertainties of the mea surement were estimated to be 5 x 10(-9) for the in-phase component an d 1,2 x 10(-8) for the quadrature component.