A METHOD OF HEAT-CAPACITY MEASUREMENT WITHOUT THERMOMETER CALIBRATION

Citation
F. Fominaya et al., A METHOD OF HEAT-CAPACITY MEASUREMENT WITHOUT THERMOMETER CALIBRATION, Review of scientific instruments, 69(1), 1998, pp. 168-171
Citations number
9
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
ISSN journal
00346748
Volume
69
Issue
1
Year of publication
1998
Pages
168 - 171
Database
ISI
SICI code
0034-6748(1998)69:1<168:AMOHMW>2.0.ZU;2-R
Abstract
A new method for measuring heat capacities that makes the calibration of thermometers obsolete is presented. The particularity of the princi ple, based on the ac steady-state method, is the insertion of a refere nce mass with a heater between the ensemble substrate+sample+thermomet er+heater, and the bath. The method could find a broad application in calorimetric devices where the thermometer must often be recalibrated due to, e.g., a drift of the thermometer's characteristics with therma l cycles or magnetic field. (C) 1998 American Institute of Physics.