LIMITATIONS OF THE MMS-10 TYMPANIC DISPLACEMENT ANALYZER

Citation
Hj. Rosingh et al., LIMITATIONS OF THE MMS-10 TYMPANIC DISPLACEMENT ANALYZER, Audiology, 37(1), 1998, pp. 1-6
Citations number
16
Categorie Soggetti
Otorhinolaryngology
Journal title
ISSN journal
00206091
Volume
37
Issue
1
Year of publication
1998
Pages
1 - 6
Database
ISI
SICI code
0020-6091(1998)37:1<1:LOTMTD>2.0.ZU;2-F
Abstract
Using the MMS-10 tympanic displacement analyser, the test-retest varia tion was determined in 20 normal hearing ears (ten subjects) for the m easurement parameters Vi and Vm. Variations were found to be on averag e 60 nl for Vi and 72 nl for Vm. The consequences of these substantial test-retest variabilities for the applicability of the device and int erpretation of the measurement results are discussed.