A COMPARISON OF KODAK EKTASPEED PLUS FILM AND THE SIEMENS SIDEXIS DIGITAL IMAGING-SYSTEM FOR CARIES DETECTION USING RECEIVER OPERATING CHARACTERISTIC ANALYSIS
Da. Tyndall et al., A COMPARISON OF KODAK EKTASPEED PLUS FILM AND THE SIEMENS SIDEXIS DIGITAL IMAGING-SYSTEM FOR CARIES DETECTION USING RECEIVER OPERATING CHARACTERISTIC ANALYSIS, Oral surgery, oral medicine, oral pathology, oral radiology and endodontics, 85(1), 1998, pp. 113-118
Citations number
27
Categorie Soggetti
Pathology,Surgery,"Dentistry,Oral Surgery & Medicine
Objective, To evaluate the accuracy of proximal caries detection compa
ring enhanced and unenhanced Siemens Sidexis CCD-based digital images
with Ektaspeed Plus films utilizing receiver operating characteristic
analysis. Study Design. Sixty extracted teeth (24 posterior and 36 ant
erior) were imaged under identical standardized geometric and exposure
conditions. Six observers, using a 5-point confidence scale, rated 12
0 proximal surfaces for the presence or absence of carious lesions by
means of three image modalities: (1) observer enhanced and (2) unenhan
ced Sidexis displays, and (3) Ektaspeed Plus films. The ground truth w
as determined by microscopic analysis of ground sections. Receiver ope
rating characteristic curves were generated with calculated areas (A(2
)) analyzed with analysis of variance for effect of reader, reading, a
nd modality. Results. Analysis of variance demonstrated significant di
fferences among readers, readings and modalities (mean square values o
f 0.012, 0.005, 0.004, F ratios of 13.604, 5.329, 5.100; p = values of
0.001, 0.043, and 0.030, respectively). Post-hoc paired comparisons o
f modalities using Tukey's statistic demonstrated that only film and e
nhanced Sidexis images were different from each other (p = 0.024). A(z
) scores were 0.7650, 0.7499, and 0.8008 respectively, for unenhanced
Sidexis, enhanced Sidexis, and Ektaspeed Plus film. Conclusion. Unenha
nced digital Sidexis images were equivalent to film for the detection
of proximal caries in this in vitro study. Observer enhanced Sidexis i
mages exhibited a statistically significant lower diagnostic accuracy
than the unenhanced digital and film images.