LOADING CAPACITY OF CFRP HIP ENDOPROSTHES ES WITH CERAMIC FEMORAL HEADS

Citation
Hj. Fruh et al., LOADING CAPACITY OF CFRP HIP ENDOPROSTHES ES WITH CERAMIC FEMORAL HEADS, Biomedizinische Technik, 42(12), 1997, pp. 369-373
Citations number
20
Journal title
ISSN journal
00135585
Volume
42
Issue
12
Year of publication
1997
Pages
369 - 373
Database
ISI
SICI code
0013-5585(1997)42:12<369:LCOCHE>2.0.ZU;2-8
Abstract
Wear of the articulating surfaces of mating components of total hip en doprostheses is the most important technical factor limiting the funct ional lifetime of the prosthesis. Biolox(R) ceramic femoral heads have been used successfully throughout the world for more than 20 years. I n vitro tests and clinical results show that the artificial modular hi p joint system employing a Caproman(R) stem and a ball made of Biolox( R) or Biolox(R)-forte is resistant to mechanical failure. The threshol d values for burst load and cyclic load prescribed by the FDA are achi eved or bettered. Clinical results support those of the in vitro tests . The femoral head made of Biolox(R) or Biolox(R)-forte used in combin ation with cups and stems made of Caproman(R) thus complies with the r equired standards.